The production of antihydrogen by the GBAR experiment at AD/ELENA requires good knowledge of the number of incident keV antiprotons, which can be problematic. We have used a commercial CMOS digital camera mounted around the experimental vacuum chamber to determine antiproton numbers from ionising particles created in the annihilation process on the surface of microchannel plate detectors which are used for beam imaging. We show that the multiplicity of emerging charged particles is as expected for individual annihilations of antiprotons with nucleons at rest, taking into account the surrounding material budget. Most of those particles are in the minimal ionising regime, but can be detected with nearly 100% efficiency in the CMOS pixel detector, while due to the thin depletion layer the device is insensitive to background gammas. Thanks to the high granularity and small pixel size millions of antiproton annihilations can be reconstructed in a dense tracking environment over a large dynamic range with good resolution. From cluster length studies of non perpendicular tracks the thickness of the depletion zone and effective detection area was estimated. The cluster length also allows for a monitoring of track angles. Antiproton numbers are determined from the number of reconstructed clusters in the CMOS sensor by means of the covered solid angle relative to a calibration measurements with well known beam intensities at the most upstream location of the GBAR apparatus. Material effects on the emerging annihilation products were estimated by Monte Carlo (Geant4) calculations, while annihilation artefacts on the complex surface of a microchannel plate are cancelled out in this approach. This method minimises largely systematic uncertainties, leading to a final error of roughly 10% for the reconstruction of absolute antiproton numbers.

Explore related subjects

Keep this discovery

BibTeXRIS

C. Regenfus, P. Adrich, I. Belosevic, F. Benkel, M. Chung, P. Cladé, P. Comini, P. Crivelli, P. Debu, A. Douillet, S. Geffroy, S. Guellati-Khelifa, P. Guichard, P. -A. Hervieux, L. Hilico, P. Indelicato, S. Jonsell, J. -P. Karr, B. Kim, S. Kim, E. -S. Kim, N. Kuroda, B. Lee, L. Liszkay, D. Lunney, G. Manfredi, B. Mansoulié, V. Martimort, M. Matusiak, V. Nesvizhevsky, F. Nez, K. Park, N. Paul, E. Perez, P. Pérez, C. Roumegou, J. -Y. Roussé, F. Schmidt-Kaler, K. Szymczyk, T. A. Tanaka, B. Tuchming, D. -P. van der Werf, D. Won, S. Wronka, P. Yzombard. 2026-09-03. Monitoring antiproton numbers with a CMOS detector in a dense-track environment. https://arxiv.org/abs/2609.04078

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-Speed Semi-FE Readout Module for ATLAS MDT at HL-LHC: Design and Production-Level Characterization

The High-Luminosity upgrade of the Large Hadron Collider (HL-LHC) introduces increased demands on the ATLAS Muon Spectrometer, particularly in terms of data throughput, timing distribution and system reliability. The Phase-II Chamber Service Module (CSM) is a key component of the upgraded Monitored Drift Tube (MDT) trigger and readout system, providing a high-speed interface between the front-end electronics and the backend systems. This paper describes the design and implementation of the Phase-II CSM, together with its validation. The results show that the CSM supports two independent optical uplinks, each operating at a line rate of 10.24 Gbps, together with clock distribution and slow control in the expected operating environment. Integration with small-diameter MDT (sMDT) chambers and tests with the prototype L0MDT trigger system are also presented. The CSM boards are now in production and will be used for installation and integration during the upcoming LHC Long Shutdown.

physics.ins-det

Spectral Discrimination of Deposited Gamma-Ray Energies in a Simulated CeBr$_3$ Scintillator

We show that wavelength measurements of individual detected optical photons may provide additional information about gamma-ray energy deposited in a CeBr$_3$ crystal when the detected-photon-count distributions overlap for nearby gamma-ray energies. Monoenergetic 662 and 629 keV gammas are used in a Geant4 simulation of a $25\times25\times20~\mathrm{mm^3}$ CeBr$_3$ crystal. Assuming a light yield of $6.0\times10^4$ photons/MeV, a wavelength-independent photon-detection efficiency of 30%, and a wavelength resolution of $\sigma_{\lambda}=40$ nm, we find that the fraction of photons reconstructed above 385 nm gives an event-level separation of $\sim$ 2 standard deviations between the 662 and 629 keV event populations selected within the same $\sim$ 1%-wide detected-photon-count interval. No timing or reconstructed interaction-position information is used. The result demonstrates, within the present simulation model, that event-dependent optical spectra can retain energy information beyond an undifferentiated photon count.

physics.ins-det

Operation of a negative ion gas time projection chamber without electronegative fill gases

The high fidelity reconstruction of particle tracks in micropatterned gaseous time projection chambers renders this technology ideal for future rare-event searches, including direction-sensitive dark matter experiments. Large drift distances are typically required for such experiments, so that the overall spatial resolution is limited by diffusion. Negative ion drift exhibits lower diffusion than electron drift and is thus an attractive option for realising a large-scale detector. The use of electronegative gases to create negative ions introduces technical challenges, most notably a reduction in gain when compared to conventional gas mixtures. In this study, we demonstrate a new method for negative ion generation via dissociative electron attachment using the conventional molecular fill gas CF$_4$. Our optical measurements of negative ion drift indicate electron attachment lengths of $<$1 mm and comparable gain to electron avalanches. The individual negative ion avalanches were also time-resolved, allowing the number of ions reaching the readout to be counted. We measure an improved energy resolution by single ion counting, relative to an integrated electron avalanche signal measured under identical gain conditions.

physics.ins-det