Time Resolution of the 4H-SiC PIN Detector
Abstract
We address the determination of the time resolution for the $\rm 100~\mu m$ 4H-SiC PIN detectors fabricated by Nanjing University (NJU). The time response to $\rm \beta$ particles from a $\rm ^{90}$Sr source is investigated for the detection of the minimum ionizing particles (MIPs). We study the influence of different reverse voltages, which correspond to different carrier velocities and device sizes, and how this correlates with the detector capacitance. We determine a time resolution $\rm (94\pm1)~ps$ for $\rm 100~\mu m$ 4H-SiC PIN detector. A fast simulation software, termed RASER (RAdiation SEmiconductoR), is developed, and validated by comparing the waveform obtained from simulated and measured data. The simulated time resolution is $\rm (73\pm 1)~ps$ after considering the intrinsic leading contributions of the detector to time resolution.
Explore related subjects
Keep this discovery
Tao Yang, Yuhang Tan, Qing Liu, Suyu Xiao, Kai Liu, Jianyong Zhang, Ryuta Kiuchi, Mei Zhao, Xiyuan Zhang, Congcong Wang, Boyue Wu, Jianing Lin, Weimin Song, Hai Lu, Xin Shi. 2021-05-31. Time Resolution of the 4H-SiC PIN Detector. https://doi.org/10.3389/fphy.2022.718071