Test of $^{116}$CdWO$_4$ and Li$_2$MoO$_4$ scintillating bolometers in the CROSS underground facility with upgraded detector suspension

Abstract

In preparation to the CROSS $2\beta$ decay experiment, we installed a new detector suspension with magnetic dumping inside a pulse-tube cryostat of a dedicated low-background facility at the LSC (Spain). The suspension was tested with two scintillating bolometers based on large-volume 116CdWO4 (CWO-enr) and Li2MoO4 (LMO) crystals. The former, a reference device, was used for testing new noise conditions and for comparing bolometric performance of an advanced Li2MoO4 crystal developed in the framework of the CLYMENE project, in view of next-generation double-beta decay experiments like CUPID. We cooled down detectors to 15 mK and achieved high performance for all tested devices. In particular both CWO-enr and LMO bolometers demonstrated the energy resolution of 6 keV FWHM for the 2.6 MeV gamma quanta, among the best for thermal detectors based on such compounds. The baseline noise resolution (FWHM) of the CWO-enr detector was improved by 2 keV, compared to the best previous measurement of this detector in the CROSS facility, while the noise of the Ge-based optical bolometer was improved by a factor 2, to 100 eV FWHM. Despite of the evident progress in the improving of noise conditions of the set-up, we see high-frequency harmonics of a pulse-tube induced noise, suggesting a noise pick-up by cabling. Another Ge light detector was assisted with the signal amplification exploiting the Neganov-Trofimov-Luke effect, which allowed to reach 20 eV FWHM noise resolution by applying 60 V electrode bias. Highly-efficient particle identification was achieved with both detectors, despite a low scintillation efficiency of the LMO material. The radiopurity level of the LMO crystal is rather high; only traces of 210Po and 226Ra were detected (0.1 mBq/kg each), while the 228Th activity is expected to be at least an order of magnitude lower, as well as a 40K activity is found to be < 6 mBq/kg.

Explore related subjects

Keep this discovery

BibTeXRIS

A. Ahmine, I. C. Bandac, A. S. Barabash, V. Berest, L. Bergé, J. M. Calvo-Mozota, P. Carniti, M. Chapellier, I. Dafinei, F. A. Danevich, T. Dixon, L. Dumoulin, F. Ferri, A. Giuliani, C. Gotti, P. Gras, D. L. Helis, A. Ianni, L. Imbert, H. Khalife, V. V. Kobychev, S. I. Konovalov, P. Loaiza, P. de Marcillac, S. Marnieros, C. A. Marrache-Kikuchi, M. Martinez, C. Nones, E. Olivieri, A. Ortiz de Solórzano, Y. Peinaud, G. Pessina, D. V. Poda, Th. Redon, Ph. Rosier, J. A. Scarpaci, V. I. Tretyak, V. I. Umatov, M. Velazquez, M. M. Zarytskyy, A. Zolotarova. 2023-07-27. Test of $^{116}$CdWO$_4$ and Li$_2$MoO$_4$ scintillating bolometers in the CROSS underground facility with upgraded detector suspension. https://doi.org/10.1088/1748-0221%2F18%2F12%2Fp12004

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-Speed Semi-FE Readout Module for ATLAS MDT at HL-LHC: Design and Production-Level Characterization

The High-Luminosity upgrade of the Large Hadron Collider (HL-LHC) introduces increased demands on the ATLAS Muon Spectrometer, particularly in terms of data throughput, timing distribution and system reliability. The Phase-II Chamber Service Module (CSM) is a key component of the upgraded Monitored Drift Tube (MDT) trigger and readout system, providing a high-speed interface between the front-end electronics and the backend systems. This paper describes the design and implementation of the Phase-II CSM, together with its validation. The results show that the CSM supports two independent optical uplinks, each operating at a line rate of 10.24 Gbps, together with clock distribution and slow control in the expected operating environment. Integration with small-diameter MDT (sMDT) chambers and tests with the prototype L0MDT trigger system are also presented. The CSM boards are now in production and will be used for installation and integration during the upcoming LHC Long Shutdown.

physics.ins-det

Spectral Discrimination of Deposited Gamma-Ray Energies in a Simulated CeBr$_3$ Scintillator

We show that wavelength measurements of individual detected optical photons may provide additional information about gamma-ray energy deposited in a CeBr$_3$ crystal when the detected-photon-count distributions overlap for nearby gamma-ray energies. Monoenergetic 662 and 629 keV gammas are used in a Geant4 simulation of a $25\times25\times20~\mathrm{mm^3}$ CeBr$_3$ crystal. Assuming a light yield of $6.0\times10^4$ photons/MeV, a wavelength-independent photon-detection efficiency of 30%, and a wavelength resolution of $\sigma_{\lambda}=40$ nm, we find that the fraction of photons reconstructed above 385 nm gives an event-level separation of $\sim$ 2 standard deviations between the 662 and 629 keV event populations selected within the same $\sim$ 1%-wide detected-photon-count interval. No timing or reconstructed interaction-position information is used. The result demonstrates, within the present simulation model, that event-dependent optical spectra can retain energy information beyond an undifferentiated photon count.

physics.ins-det

Operation of a negative ion gas time projection chamber without electronegative fill gases

The high fidelity reconstruction of particle tracks in micropatterned gaseous time projection chambers renders this technology ideal for future rare-event searches, including direction-sensitive dark matter experiments. Large drift distances are typically required for such experiments, so that the overall spatial resolution is limited by diffusion. Negative ion drift exhibits lower diffusion than electron drift and is thus an attractive option for realising a large-scale detector. The use of electronegative gases to create negative ions introduces technical challenges, most notably a reduction in gain when compared to conventional gas mixtures. In this study, we demonstrate a new method for negative ion generation via dissociative electron attachment using the conventional molecular fill gas CF$_4$. Our optical measurements of negative ion drift indicate electron attachment lengths of $<$1 mm and comparable gain to electron avalanches. The individual negative ion avalanches were also time-resolved, allowing the number of ions reaching the readout to be counted. We measure an improved energy resolution by single ion counting, relative to an integrated electron avalanche signal measured under identical gain conditions.

physics.ins-det